During the 1980s, considerable effort was made in
developing olfactometric odour measurement techniques in the Netherlands
and elsewhere in Europe. Initially, the application of these olfactometric
results was limited mainly to comparing odour emissions from various
manure treatment systems in intensive animal production. In 1985,
the Victoria EPA in Australia first introduced legislation based
on olfactometer results and air dispersion model (Ausplume). In
North America, despite earlier interest in olfactometric measurement
techniques during the 1970s, it was not until the mid 1990s that
North American universities set up olfactometry laboratories to
investigate odour from animal production.
The development of olfactometric measurement techniques
continued in Europe and resulted in the introduction of the first
draft European Standard for odour measurement by dynamic olfactometer.
In Australia, a national workshop on odour measurement standardization
was held in 1997 and consensus was reached to adopt the draft
European standard. In 2001, Australia published the first official
standard ahead of European countries. European countries have
officially agreed to adopt the CEN standard in early 2003.
In summary, most olfactometers currently used around
the world can be categorised in three groups on the basis of how
the dilution is achieved:
• static method (syringe method in USA,
triangle bags in Japan);
• rotameter/fixed orifice based olfactometers (VIC EPA
B2 in Australia, TO8 in Germany, IITRI in USA) and
• Mass Flow Controller (MFC) based olfactometers (Ac’scent
olfactometer in USA, Olfaktomat as used in the Netherlands).
As early as 1960, the syringe was used as an olfactometer
to prepare the odour sample and was inserted into the panellist’s
nostril for evaluation. The method was later published in 1978
by the American Society for Testing and Materials. The concentration
ascending presentation order, the sample presentation flow rates,
the sample losses and cross contamination were widely questioned
in the early 80’s. The method has finally withdrawn in 1986.
In 1972, Japan improved the method to replace the syringe with
a 3-litres bag and has published the Triangle Odor Bag Method.
The method has now been widely adopted in Asian countries until
today. The disadvantage of the method is the manual operation
which requires a long time to prepare the 15- 18 bags and to calculate
the results. The accuracy and repeatability of the results are
very poor. However, the use of logarithm results can improve the
accuracy and repeatability but this has little implication in
a real application. The delivery of the odour samples is still
achieved by a syringe. This has cast great doubt on how much improvements
the Triangle Odor Bag method has made.
Rotameter based olfactometers are currently used
in many laboratories in Australia and elsewhere. The rotameters
are extremely sensitive to downstream pressure variations that
could result in errors in rotameter readings of up to 25%. Such
pressure variations may be occurred during the mixing of clean
air and odorous air to create the required dilution ratio or subsequently
during the sample presentation of the diluted sample. The latter
may be accentuated by the use of an enclosed sniffing mask, adversely
affecting overall performance of the olfactometer. The manual
mode of operation for rotameter based olfactometers makes it impossible
to meet stringent instrumental performance criteria, particularly
at the high dilution ratio end of the range. Furthermore, high
labour costs when using manual data input for monitoring panellist
performance and in the data processing used for retrospective
screening may also make compliance with the proposed standard
excessively expensive.
MFC based olfactometers have automated the diltuion
process but it has yet to demonstrate instrumental performance
over a period of time. MFC based olfactometers are also sensitive
to the downstream pressure of the flow measurement devices. The
backpressure occurring during mixing can be compensated for by
instrumental calibration. However, pressure variations occurring
during the sample presentation stage cannot be predicted and therefore
cannot be compensated for by calibration. Backpressures can vary
from panellist to panellist. In practice, the reduced flow arising
from the specific personal characteristics of a panellist will
be sensed by the mass flow meter resulting in the valve being
further opened. However, the presentation time for each panellist
is long (10 – 30 seconds) in comparison with the response
time of the mass flow controller to change the valves (several
seconds). These unstable conditions will be repeated many times
during the session. As a result, the actual dilutions of odour
samples at the sniffing ports can be highly variable.
Furthermore, the MFCs are susceptible to contamination
buildup that can alter the calibration and result in the reduced
performance. The tiny space between the temperature elements inside
the mass flow meter can be easily contaminated or blocked. The
MFC is really designed for single component gas and better suited
to a clean and non-sticky gas. In particular, the odour samples
can sometime be very sticky and dusty. Therefore, MFC based olfactometers
could easily suffer from the poor performance of the MFCs during
the operation. This has proved to be a major limitation in the
use of MFC based olfactometers. Flushing the MFCs may take hours
and is not effective at all. Over a period of usage, dust and
residuals eventually become irreversibly adhered to the surfaces
of the temperature elements and the MFC must be replaced. The
performance of MFC based olfactometer in delivering the required
dilution ratio cannot be guaranteed.